Built in Medical Isolator MI 1005 E
TÜV certified isolator for galvanic network isolation of medical electrical devices
Safety: IEC 60601-1:2005/AMD1:2012/AMD2:2020 (Edition 3.2)
EMV: EN 60601-1-2:2015 (IEC 60601-1-2:2014) (Medical electrical equipment)
In addition to network isolators with fixed cables, the medical isolator MI 1005 E offers a simple and cost-effective way to isolate medical devices, starting at the wall of the housing. This makes the insulator particularly well-suited for medical device makers. Using standard Harting PushPull technology, the MI 1005 E Medical Built In insulator can be installed quickly and easily into existing devices. Long-term availability also offers planning certainty necessary for new product development.
Under stress conditions the MI 1005 isolator was tested up to a isolation stength of 10 kV DC (10 Sec.) and 5 kV AC (60 Sec.). Furthermore, 8 kV contact discharge and 16 kV air discharge are easily fulfiled.
Benefits:
- Ideal for medical device manufacturers
- Insulation from the device housing wall
- Easy installation thanks to Harting PushPull technology
- Offers planning security – long-term availability
- Passive component – without additional power supply
Specification
| Important data | |
|---|---|
| Case color | Light gray |
| Dimensions (DWH) | 65x29x23 mm |
| Weight approx | 50 g |
| Operation mode | Continuous operation |
| Dielectric strength signal and shielding | 5000 V 50/60 Hz for 1 min. and 10000 V 50/60 Hz for 10 s. |
| Data throughput | 10/100/1000 MBit/s |
| Ports | 2x RJ45 IN/OUT |
| DC Stress Test (10 Sec) | 10 kV |
| Network Specifications | IEEE 802.3ab 10/100/1000-BaseT; Twisted-Pair; auto-conf (behaves completely transparent in the network) |
| Material | UL94V-0 / RoHS conform / REACH |
| Produkt Rating (UL) | Passive, electronic components, insulation DI (250 V AC / 300 V DC) tested at 5 kV AC |
| Operating time between failures | 0,21 x 10 ^ 8 (1 failures in 21 000 000 h) |
| Return loss | Min. 10 dB |
| Insertion loss | Max. 1,1 dB |
| Standards | IEC 60601-1 (3rd Edition); IEC 60601-1-2; UL 60601 (UL Listed E504923-D1000); IEEE 802.3ab |
| Temperature | -10 °C up to +70 °C (operation/ storage/ transport) |
| Relative air humidity | 10 % up to 90 % (non-condensing (operation/ storage/ transport) |
| Air pressure | 700 hPa up to 1060 hPa (operation/ storage/ transport) |
| Warranty | 5 years |
| Scope of delivery | Netzwork Isolator MED MI 1005 |
| Item No. | 2005674 |
Login