- Order number: 2011786
The only multiple socket outlet that secures your system
according to the DIN EN60601-1 standard.
When combining medical electrical devices (ME devices) and medical electrical systems (ME systems), the use of a multiple socket is essential. Chapter 16.9.2.1 of the standard for electrical safety of medical electrical equipment stipulates that multiple sockets must be supplied via an isolating transformer. The multiple socket outlet with galvanic isolation, consisting of Baaske MED X socket strip and Baaske MEX T1200, meets the requirements of the standard DIN EN 60601-1 Chapter 16.9.2.1 and thus offers a maximum of safety and standard conformity of the system.
Technical Data multipla socket isolation transformer MEDX 1200VA
Customized production
Discover the versatility of this system, which can be optimally adapted to your requirements. Up to 8 modules can be ordered using the modular principle, which can optionally be manufactured into a socket strip with up to five metres of cable. The concept of modular socket outlets offers a cost-efficient solution that allows you to procure only the required number of socket outlets. Future expansions of your electrical system do not require a new product. The missing module can easily be retrofitted by us.
Minimum order quantity=50 pieces
Multiple socket incl. 1200 VA isolating transformer
The European standard EN60601-1 for the electrical safety of medical electrical equipment stipulates in Chapter 16.9.2.1 that multiple sockets must be supplied via an isolating transformer. This multiple socket for ME systems was specially developed on the basis of this standard and is currently the only multiple socket with galvanic isolation. Consisting of the Baaske MED X socket strip and the Baaske MEX T1200 isolating transformer, this system meets the requirements of the standard DIN EN 60601-1. The integrated 1200VA isolating transformer was specially developed for the power supply of supply-critical medical electrical systems (ME systems) in hospitals and medical practices and has a high reliability. In addition, this system protects your patients, your equipment and also itself from excessive inrush currents, half-wave failures, overheating and excessive voltages. The special front interface for the integration of an insulation monitor offers even more safety. Any insulation faults that may occur can thus be safely avoided.
A system that offers maximum safety and conforms to the medical standards.
Solid aluminium multiple socket
with galvanic isolation
A tailor-made solution for your systems. A power strip that adapts ideally to the requirements of medical electrical devices (ME devices). The unique design allows individual production without sacrificing important features. Up to 8 modules can be ordered as a modular system, which can be manufactured with a connecting cable of up to five metres. Each module has an additional equipotential bonding (ZPA) and two G-melting inserts each. The system-integrated isolating transformer with 1200VA additionally offers sufficient capacity to optimally protect your electrical system against malfunctions. The potential equalisation (ZPA) on the multiple socket outlet housing offers a further plus in safety; it enables the strip to be used simultaneously as a local ZPA distributor.
Convince yourself of the versatility of this system, which adapts optimally to your requirements and conforms to standards.
Features
Isolating transformer and multiple socket in one device
Electrically safe galvanically isolated
1200VA (230V primary/ secondary, 50/60 Hz)
incl. trigger protection(s)
Customizable
2Double fuses per socket insert
Additional equipotential bonding (EPB) on the housing
Electronic inrush current limitation
Interface for insulation monitor
Temperature switch primary
Half wave failure detection
11mm strong feed cable end and kink protection at the housing
Technical Data multiple socket: | |
---|---|
Fuse | per socket insert 2 G-fuse-links, each 5x 20mm |
EPB connector | according to DIN 42801 |
Load per socket | max. 2000W |
Total load power strip | 3600 W |
Technical Data isolation transformer: | |
---|---|
Power output | 1200 VA max. |
Power consumption | 1200 VA max. |
output sockets | 1x IEC 320 |
Protection class | IP67 |
Minimum order quantity | 50 pieces |